null
Savillex

C100s PFA Nebulizer with Carbon Probe, 100 cm, ICP-MS Sample Introduction, each

Part Number:
SVX-800-4-010-04-00
Lead Time:
4 weeks
Shipping:
FREE SHIPPING on most orders over $50*
Quantity:
each
Size:
70 cm
Material:
PFA
Country of Origin:
Made in the USA
Price: $1,943.69
Frequently bought together:

Description

The C100s Nebulizer with Carbon Probe (100 cm) is engineered for ultra-high sensitivity in normal and cool plasma modes, specifically tailored for semiconductor applications.

Features:

  • Microconcentric PFA s-type nebulizer
  • Demountable PFA uptake line with unique zero dead volume connection, ensuring the same washout performance as an integrated uptake line
  • Robust design allows for backflushing, eliminating the need for additional cleaning tools
  • 100 cm uptake line, consisting of approximately 85 cm of flexible uptake line and 15 cm carbon probe
  • Encapsulated carbon probe with 2.9 mm OD and 14.4 cm length, including encapsulation
  • Compatible with ESI SC Micro autosampler and CETAC ASX 110/112 with new style arm (from Feb 2015)
  • 6 mm OD body with 4 mm gas port
  • Free aspirates at 100 µl/min +/-20 µl/min, or can be pumped
  • Can use with supplied uptake line or connect directly to line from ESI FAST systems
  • Gas line connection kit included. For connection to the PE NexION 2000, order Part #830-052

This nebulizer supersedes Part #800-1-010-04-00.

Note: This nebulizer is designed for use only with standard sample introduction systems (ambient temp or chilled spray chamber). Where a desolvator is used, a nebulizer type “Nebulizer for Desolvators” must be selected.

Specifications:

  • UPTAKE RATE µl/min: 100
  • AUTOSAMPLER COMPATIBILITY: ESI SC MICRO/CETAC ASX110/112
  • PROBE MATERIAL: CARBON FIBER
  • FOR DESOLVATOR USE: NO
  • A Inches (nn): 3.53" (89.8)
  • B Inches (nn): 0.24" (6.0)
  • C Inches (nn): 5.67" (144.0)
  • D Inches (nn): 0.11" (2.9)
  • TUBING LENGTH Inches (nn): 39.37" (1000.0)

Additional Details:

  • Robust - can be backflushed; no cleaning tools needed
  • Body fits any spray chamber port (Scott-type, cyclonic or desolvator) that accepts a 6 mm OD nebulizer - choose the regular or desolvator version to match your system
  • Installation: tighten the spray chamber ferrule nut until it contacts the nebulizer body, then a further 1/8 turn only - overtightening can deform the body and reduce sensitivity
  • Always handle the nebulizer by the body; never grip or pull on the uptake line when inserting or removing it
  • To clear stop-flow caused by air gaps, backflush: with nebulizer gas on, lightly place a gloved finger over the nebulizer tip
  • For ultratrace (e.g. semiconductor) work, aspirate fresh high purity 5% HNO3/2% HF for at least 8 hours or overnight before first installation; do not recirculate the acid
  • Gas line connection kit includes 1 m of 4 mm OD PTFE gas line, a Festo 4 mm/6 mm push-fit connector and a barbed connector for Tygon gas lines; additional kits are part 830-051
  • s-type inner body design gives about 10% higher sensitivity in normal plasma and 15% higher in cool plasma than previous-generation C-Flows
  • Uptake line and capillary are PFA, resistant even to concentrated TMAH and mineral acids; no measurable elemental background contribution
  • Uptake line connector nut carries a groove pattern identifying the uptake rate of the line
  • Ships with a checkout certificate showing sample uptake rate across 0.6-1.1 SLPM argon nebulizer gas flow
  • Uptake rate specification ±20% (80-120 µL/min) at 0.7 SLPM nebulizer gas flow
  • Designed to operate over 0.6-1.1 SLPM argon nebulizer gas flow; 0.7-1.0 SLPM is typical on ICP-MS
View AllClose

0 Reviews

View AllClose
MPN:
800-4-010-04-00